Journal of Chemical and Pharmaceutical Research (ISSN : 0975-7384)

header
Reach Us reach to JOCPR whatsapp-JOCPR +44 1625708989
All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.

Original Articles: 2014 Vol: 6 Issue: 1

A simple HPLC-UV method for the determination of clindamycin in human plasma

Abstract

This study describes a simple high performance liquid chromatographic (HPLC) method for the determination of clindamycin in plasma. Analysis was carried out using a Varian® Pro Star HPLC unit equipped with an online degasser. A reversed-phase ACE® C18 column of dimensions 250x4.6mm, particle size 5μm was used. The mobile phase was made up of 0.02M disodiumhydrogen phosphate buffer (pH of 2.9) and acetonitrile at a ratio of 71:29 v/v, running through the column at a flow rate of 1.5ml/min and with ultraviolet (UV) detection set at a wavelength of 195nm. Clindamycin was separated from plasma proteins by protein precipitation with ice cold acetonitrile. Clindamycin and the internal standard phenobarbitone eluted after 3.96 and 7 minutes respectively. The method was validated for linearity in the working concentration range of 0.5-20μg/ml. Linearity was observed with a coefficient of determination (r2) of 0.990. The recoveries obtained were all above 82% and the limit of quantification and limit of detection were 0.2μg/ml and 0.1μg/ml respectively.

http://sacs17.amberton.edu/